System Information
System User Guide
D5033331 A
AB SCIEX TripleTOF
®
5600/5600+ Instruments
23 of 116
5. Click Configure Pump.
Figure 1-4 Harvard Syringe dialog
6. Select the COM port for the connection to the syringe pump.
7. Click OK until the Hardware Configuration Editor dialog box appears.
8. Activate the hardware profile.
Adjust the Integrated Syringe Pump Position
1. Press the Release button on the right side of the syringe pump to lower the base and
then insert the syringe as shown in Figure 1-5.
Make sure that the end of the syringe is flush against the base and that the shaft of
the syringe rests in the cutout.