252
SPT
Item Description
Injector
Coolant
On or Off
Turns the coolant valve on when the SPT injection device is cooling
down. The cryogenic will not be used if the Off position is selected.
Enable
Coolant at (C)
30-450 ºC
Specifies the temperature for the coolant valve to turns on.
Coolant
Timeout (min)
0.01-999.99 min
Safety measure to save coolant after the specified time if either the
SPT fails to reach its set temperature or the GC does not go into run.
Temp (C) -185-450 ºC
Hold (min) 0.01-999.99 min
Total (min) 0.01-999.99 min
Cannot be edited.
Add Adds a line to the spreadsheet.
Insert Inserts a line above the currently selected row in the spreadsheet.
Delete Deletes the currently selected row(s) in the spreadsheet.