Item Description
Edit/Lock Calibration
Data...
Opens the Coefficients window to edit calibration
coefficients for the Method.
Coefficients Table
Item Description
Retention Time Shows the retention time of each peak.
Peak Name Shows the name of each peak.
Lock Coeffs. checked or not checked
If checked, coefficients will not be updated on calibration
runs. Prevents specific calibration coefficients from
changing on subsequent calibration runs. Can be used to
manually enter calibration coefficients for peaks that will
not be identified in the calibration standard.
coefficients: X
3
, X
2,
X,
intercept
-100,000,000.0 to 1,000,000
Sets the coefficients that define the calibration curve for
each peak.