Axis BTA 2100D Manual del operador Pagina 99

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 597
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 98
97
scanned out of the ion trap. Thus, when the multiplier gain is changed, there is
no change in the actual quantity of ions formed and measured in the prescan
measurement, but the peak area of the AGC peak will be changed. The AGC
software will misinterpret this change in peak area as a real change in the
number of ions created in the ion trap and will adjust the ionization time
accordingly. An incorrect ionization time may cause a change in mass resolution.
The AGC software is dependent on the multiplier gain remaining constant in
order to measure the quantity of ions in the ion trap correctly and set the proper
ionization time.
Under certain circumstances (low background, small signal) increasing the
electron multiplier gain by +100 or +200 volts above the 10
5
gain can decrease
the detection limit. As explained in the note above, the same multiplier gain must
be used for determining the calibration curve and analyzing unknown samples.
Axial Modulation
The number of ions stored in the ion trap cavity is proportional to the ionization
time and the filament emission current. Although the maximum ion-storage
capacity of the ion trap is approximately 10
6
to 10
7
ions, space-charge repulsion
occurs when the number of ions exceeds 10
4
to 10
5
. Space-charge interactions
cause a loss in unit mass resolution. To overcome this limitation, a small ac
voltage, the axial modulation voltage, is applied to the endcap electrodes.
The axial modulation voltage is applied at a fixed frequency and amplitude during
the ramp of the rf voltage. The frequency of the axial modulation voltage is 485
kHz, which is about one-half of the frequency of the rf voltage. Only when an ion
is about to be ejected from the ion trap cavity is it in resonance with this
frequency. When an ion comes into resonance with this frequency, it moves
away from the center of the trap, where the field generated by the rf voltage is
zero (and space-charge effects are strong), into a region where the field
produced by the rf voltage is strong (and space-charge effects are small). As a
result, the ejection of the ion is facilitated, and mass resolution is significantly
improved. The default axial modulation voltage is 4.0. Once the optimum axial
modulate ion amplitude has been set for a given instrument, it need not be
changed.
Emission Current
The number of ions formed in the trap is a function of both the ionization time
and the intensity of the electron beam. In general, increasing the filament
emission current increases the quantity of ions produced. The relationship
between the filament emission current and the measured TIC (i.e., the number of
ions generated in the ion trap) is not a strictly linear one. In particular, the
emission current and the measured TIC do not vary linearly for emission current
values above 40 µA and at high sample concentration.
In certain cases, a high emission current can be used to increase the sensitivity.
One of these is EI/MS/MS where the MS/MS isolation step eliminates the higher
background that would normally arise from a high emission current.
Ion Trap Temperature
After the system has been shut down and the ion trap vented to the atmosphere,
we recommend an initial trap temperature of 220°C for the rapid removal of water
vapor. After this period, a reduction in trap temperature to 150-175°C will extend
the turbomolecular vacuum pump lifetime. Some applications may require higher
Vista de pagina 98
1 2 ... 94 95 96 97 98 99 100 101 102 103 104 ... 596 597

Comentarios a estos manuales

Sin comentarios