Item Description
Fill Volume (uL) 0 to 10.0 microliters with 10 uL syringe
0 to 5.0 microliters with 5 uL syringe
Specifies the sample volume that will be used for each
fill stroke.
Fill Strokes 0 to 99
Specifies the number of the times the sample will be
“pumped” in and out of the syringe before loading the
sample volume into the syringe.
Sample Air Gap Yes or No
If Yes, 1 microliter of room air will be drawn into the
syringe to create an air gap before the sample plug.
Air Plug after Sample 0 to 10.0 microliters with 10 uL syringe
0 to 5.0 microliters with 5 uL syringe
Specifies the volume of room air that will be drawn into
the syringe after it is loaded with sample.
Viscosity Settings
Item Description
Viscosity Delay (sec) 0 to 9.9 seconds
Specifies how long the syringe is to remain in the
sample vial after drawing up the sample.
Fill Speed (uL/sec) 0.1 to 50.0 microliters per second with 10 uL syringe
0.1 to 25.0 microliters per second with 5 uL syringe
Specifies the speed at which the sample will be drawn
up into the syringe.
Inject Speed (uL/sec) 0.1 to 50.0 microliters per second with 10 uL syringe
0.1 to 25.0 microliters per second with 5 uL syringe
Specifies the speed at which the contents of the
syringe will be expelled into the injector.
Pre-Inj Delay (sec) 0 to 99.9 seconds
Specifies the length of time the syringe needle resides
in the injector before expelling the syringe contents.
Post-Inj Delay (sec) 0 to 99.9 seconds
Specifies the length of time the syringe needle remains
in the injector after expelling the syringe contents.