Moving Peak Start and End Points
Peak start and end points are indicated on the chromatogram as lines or
triangles (depending on the options selected in the Preferences dialog box). You
may change the position of these points by dragging them with the mouse.
Right-click on a point to return it to its original position or view the peak event
information.
Manually positioned points are drawn as solid triangles.
Drag a point to change its position.
When you perform reintegration on a chromatogram with manually positioned
peak events, you are asked whether you wish to include the changes when
calculating results.
Use the manually placed peak events and reintegrate.
Discard the manually placed peak events and reintegrate.
Abort reintegration.
If you use the manually positioned peak events, peaks whose areas are affected
by the event will be flagged with a “U” (user-modified) in the results report.
Calculating Results
Results can be calculated for all opened chromatogram channels using the Data
Handling parameters in the active Method. Open the Reintegration list by clicking
on the button in the Main toolbar, or by selecting it from the Results menu.
Open the Reintegration List.