Type 3 EFC (for 1041/1061 Injectors)
If you indicate that a Type 3 EFC is installed, a flow ramp spreadsheet will
appear.
The first row of the spreadsheet contains the Initial flow and hold time for the
Type 3 EFC. The EFC will equilibrate to this setting when the method is
activated, and will be restored to this setting when each chromatographic run is
completed.
Rows after the first row of the spreadsheet contain the programmed settings for
each of the ramp segments. Each segment will ramp to the specified flow at the
specified rate (assuming the rate is achievable), and then hold the flow for the
specified time. (Note that the Rate in the first row is always blank and cannot be
edited. Also note that the entire Total column cannot be edited.)