453
Item Description
Save 2 Saves overlay curve 2 as the curve for the peak.
Point Info Dialog Box
Item Description
Exclude Selected Point from
Calculation
If checked, marks the selected point as an outlier,
excluding it from calibration calculations. Excluded
points are never lost, and can always be included
again at a later time.
Amount Shows the amount of the standard represented by
the selected data point.
Peak Size Shows the size of the peak represented by the
selected data point.
Deviation Shows, as a percentage, the deviation of the
selected data point from the curve. Used to see
how far off the selected point is from the result
expected from the curve.
Data File Shows the name of the Data File represented by
the selected point.
Level Specifies the level from which to select a point.
Replicate Specifies the replicate representing the point.
Select a point among those in the selected level.
Next Select the point at the next highest level on the
curve.
Previous Select the point at the next lowest level on the
curve.
Coefficients Dialog Box
Item Description
coefficients: X
3
, X
2,
X,
intercept
0.000000 to 1,000,000
Sets the coefficients for each peak.
Restore Restores coefficients to their last saved values.