Item Description
splitless runs in which the split vent is opened.
Injector EFC, Gas Saver Flow The lowest system flow rate that will be
maintained during the gas saver period. This
has a range of 1-100 mL/min and defaults to
20 mL/min.
Injector EFC, Gas Saver Timeout A non-zero value entered for this parameter
will cause the system to initiate a timer once a
method is activated. If no system actions such
as a run start or another method activation
occurs before the Gas Saver Timeout is
reached, then the system will reduce the flow
through the system to the Gas Saver Flow. A
method activation action will clear the Gas
Saver Flow and bring the system to Ready for
subsequent injections. A zero value entered
for the Gas Saver Timeout will cause the Gas
Saver Flow to be ignored and no Gas Saver
actions will be performed.
Injector EFC, Auto Calibrate EFC
Flow
Automatically calibrate the EFC system.
When it is finished, the screen will display
either a successful completion or detected
problem message. No other GC functions can
be performed during AutoCalibration. Note
that AutoCalibration should be done on a
semiannual basis.
Injector EFC, Calibrate Septum
Purge Flow
Calibrate the septum purge. After
installing a new column, the septum
purge should be calibrated. Set the
desired column head pressure. Then
measure the actual septum purge flow
rate and enter this value in the septum
purge field. The septum purge flow rate is
adjustable but is typically set at 3 - 5
mL/min.
Septum Purge Calibration, Desired
Head Pressure
The desired column head pressure in psi.
Range is 0.1 - 100 psi.
Septum Purge Calibration,
Measured Flow Rate
The septum purge flow rate is adjustable but
is typically set at 3 - 5 mL/min.
Print Actual Sends the 3900 setup parameters to the
default printer.
Close & Update Saves the changed parameters and
reinitialize the 3900 with the new parameters.
AutoSampler, Type None, 8400, 8410
Displays the type of autosampler that is
connected to the 3900 GC.
AutoSampler, Syringe Volume
5 L, 10 L and 100 L.
Enter the syringe volume you will install and
use with your CP-8400/8410. The 3900 will
use this value to set the correct entry
selections and any needed internal settings to
correctly use the indicated syringe size.
Incorrectly setting this parameter may cause