290
Micro-TCD Detector
If you indicate that a Micro-TCD Detector is installed, an Oven Power switch and
an Electronics switch will appear in the top portion of the window, and additional
detector adjustments will appear below the spreadsheet.
NOTE: The Micro-TCD cell temperature is set to 110 C and is not adjustable.
Indicate that a Micro-TCD Detector is installed
Indicate whether or not you will use the installed Micro-TCD Detector by clicking
on the “On” or “Off” radio buttons.
You can time-program the detector range, autozero, and polarity in the
spreadsheet.
Select the desired TCD Time Constant by clicking on the “Slow” or “Fast” radio
buttons.
Select the desired Carrier Gas from the available choices.
The first row of the spreadsheet contains the Initial Range, Autozero, and
Polarity for the Micro-TCD Detector. The Micro-TCD will switch to this setting
when the method is activated, and will be restored to this setting when each
chromatographic run is completed.
Rows after the first row of the spreadsheet contain the time-programmed settings
for the Range, Autozero, and Polarity. They will be switched to the value in each
row at the indicated time.